MAIN TECHNICAL PARAMETER
Single bed with single bulb
Rotary anode X-ray tube unit and tangential annular tubes
Single-phase full-wave rectification high-voltage generator
Power voltage (V) photograph kilovolt (KV) and infinitely variable control and electric mechanism
Be equipped with the Monistat for the filament of X-ray tube and space complementary
Photographic volume KV, mA and S subsection grading and interlock protection
Adopt the digital circuit timer Grading according to R10 priority coefficient ,which is exact in time control
High-voltage primary uses the zero controlled circuit of silicon-controlled rectifier of large power
Photographic bed can move in length and breadth
The photographic bed, pillar ang vibrating ray-filter are in a whole without top and bottom track
General Photography
Filter-ray device photography
ITEM
|
CONTENT
|
INDEX
|
Demand of power supply
|
Capacity
|
≈ 30 KVA / two –phase 30 KVA
|
Voltage frequency
|
380 V/220V + 10% 50 HZ + 0.5 HZ
|
Internal resistance
|
0.75 W (380) 0.25 W (220V)/0.75W at 380V;0.25W at 220V
|
Power supply
|
Adjustable range
|
380V, 380+10%220V,220+10%,380V+10% at380V;220V+10% at 220V being adjustable continuously
|
Photograpy
|
Voltage
|
50~100KV *50 ~125 KV
50~100KV *50 ~125 KV being adjustable continuously
|
Current
|
50 mA 100 mA 50 mA 100 mA 200 mA 300 mA
Small fucos : 50 mA, 100 mA big focus 50 mA, 100 mA 200 mA, 300 mA
|
Time
|
0.04~6.3s 23 0.04~6.3s 23 grades together grading according to R10 coefficient
|
High-voltage generator
|
Capacity
|
20 KVA / 20 KVA (instantaneous)
|
Max.Dc output voltage
|
100 KV * 125 KV
|
Max.Dc output current
|
200 mA
|
X-ray tube
|
model
|
XD51-20, 40/100 * XD51-20,40/125
|
focus
|
2 X 2/small focus 1 x 1 mm ;big focus : 2x 2 mm
|
|
Length XwidthXheigt of bed face
|
2000 x 710 x 700 mm
|
|
Moving range of bed face
|
680 mm 20 mm / 680 mm in length : 200 mm in breadth
|
(YZ-SCIII) Photographic bed
|
Ray – Filter of bed
|
> 500 mm / travel in length > 500 mm
|
(YZ-SCIII)
|
N28 / grid density N28
|
|
F0 100 cm / convergence distance fo 100 cm
|
|
R8/grid eatio r8
|
Pillar of X-ray source unit
|
Moving in length along photographic bed
|
1800 mm 1600 nn / 1800 mm or 1600 mm
|
Distabce to ground as moving up and down (along upright post)
|
650 ~ 1800 mm
|
Rotary around the center of cross arm
|
+ 180 0
|
Rotary arount the shaft line of x-ray annular tubes
|
10 0 ~ 60 0 ~ + 120 0
|
Max size cassette for photograph
|
356 x 432 mm (14” x 17”
|
|
|
Reviews: